GRATINGS +1 435.752.4301 SPECTROMETERS +1 919.544.7785 OCT +1 919.544.7785
Upgrade your current Raman setup with this sensitive standalone spectrometer, or build a new system from the ground up. Either way, you’ll have maximum configurability in choice of laser and sampling optics, and control over resolution and signal level with our user-interchangeable slit.
DETECTOR COOLING OPTIONS | ||||
Ambient (-A) | TEC-Regulated (-R) | TEC-Cooled (-C) | ||
Spectral Range* | 275-2000 cm-1 | 275-2000 cm-1 | 275-2000 cm-1 | |
Resolution | 15 µm slit | 5 cm-1 | 5 cm-1 | 5 cm-1 |
25 µm slit | 7 cm-1 | 7 cm-1 | 7 cm-1 | |
50 µm slit | 13 cm-1 | 13 cm-1 | 13 cm-1 | |
Detector TEC setpoint | None | 10 ± 0.2°C | -15 ± 0.2°C | |
Integration time | 3 ms – 60 s | 3 ms – 60 s | 25 ms – 60 s | |
Laser Type | 633 nm, single mode | |||
Laser Power | Up to 50 mW, control via ENLIGHTENTM software | |||
Working distance (from face of lens) | 22 mm | |||
Laser spot size (nominal) | 60 µm |
*Start and end wavenumber may be customized, but total range is fixed. Contact us for options.
Note: Specifications are subject to change based on available components & manufacturing data.
We offer three detector cooling options for the WP 633 Raman spectrometer, allowing you to balance your desired signal to noise and temperature stability against power draw and cost for maximum value:
Need advice? Contact us to discuss what’s best for your specific application.
Operating Software | ENLIGHTENTM Raman spectroscopy software & SDKs included |
Library Matching | Identify Raman spectra via Wiley's KnowItAll Raman Identification Pro (free trial, fully integrated into ENLIGHTENTM) |
Data Interface | USB 2.0 Type B receptacle |
Max Sampling Rate | > 250 Hz |
Operating Temperature | 0°C to 40°C, non-condensing |
Dimensions | 16.5 x 16.2 x 6.7 cm |
Weight | 1.8 kg |
We’ve made it our mission to design compact, portable Raman without compromise – instruments that deliver superior sensitivity at a cost-effective price point. We place one of our own perfectly matched, patented volume phase holographic (VPH) gratings at the heart of every spectrometer. With uniform response and low loss, these gratings enable a compact and highly efficient all-transmission optical design that minimizes loss and maximizes stability for consistent, high-SNR spectra.
The pattern of peaks in a Raman spectrum is unique to the sample, providing a fingerprint which can be used to identify or analyze the material. The ‘fingerprint’ region up to 1500 cm-1 is often enough to confirm identity, while the extended ‘functional’ region out to 3600 cm-1 includes bands for additional, functional groups in a molecule. In addition to our standard models, we can design custom range spectrometers for OEMs. See below for common Raman spectral bands.
Coming soon
Need something slightly different? We offer a variety of alternate spectrometer models and system setups, and can offer expert advice on which may be the right match for your application. Consider the option below, or explore our full family of Raman products.
No OEM models are available for the WP 633 Raman spectrometer at this time.
Choosing the right Raman wavelength for your application is a question of balancing sensitivity and selectivity. Although shorter wavelengths deliver the strongest Raman signal, fluorescence background can degrade signal-to-noise ratio (SNR) if the wrong wavelength is used. That’s why we offer wavelengths from 248-1064 nm, matching your specific needs and sample type to the best wavelength for the application. Explore our alternate wavelengths below. Still unsure of what you need? Contact us for a personal consultation or sample testing.
Content Coming Soon
Content Coming Soon!